4-24
Chapter 4 Function Settings
4
1 Press the
ENT
key for calibration
setting.
The AREA1 calibration display appears.
Press the
FUNC
key to make AREA2
settings.
2 Input T1-A Value.
PROG No.
PRINT
AREA DISP
ESCG
C
ENT
HOLD
ZERO
FOCUS
A
POSITION
HI
GO
LO
Press these keys to set
the value.
The input part
flashes.
Indicates that
the T1-A value
is displayed.
Indicates that AREA1
is selected.
Press this key to set
the value to a
positive or negative
one.
3 Press the
ENT
key.
The input value is entered and the value in the next line flashes.
4 Input T1-B Value.
5 Press the
ENT
key.
6 Input T2-A Value.
7 Press the
ENT
key.
The input value is entered and the value in the next line flashes.
8 Input T2-B Value.
9 Press the
ENT
key.
The input item is entered and the calibration setting display appears again.
HI
GO
LO
HI
GO
LO
The input
part flashes.
Note: If the calibration conditions are not satisfied, “Error-1” will appear on the main
display. Reset the error status by pressing the
ENT
key or
ESC
key and make
the calibration settings again.