Keyence LS-3100/3100W Series Bedienungsanleitung Seite 86

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 104
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 85
80
If any conditions occur that could be considered as equipment failure or
malfunction, follow the inspection procedures given below.
CHAPTER 9
TROUBLESHOOTING GUIDE
Check
item
Problem
Details of inspection Corrective action
Reference
page
Measure-
ment
value
display
Measured value is not
displayed.
Nothing appears on the
display.
Is the power switch ON?
Is the AC power cable con-
nected?
Turn ON the switch.
Connect the AC power cable.
p. 5
"------" is displayed. Are targets in position?
Is the entire scanning area
completely covered?
Is the appropriate measuring
segment selected?
Is the scanning head properly
connected?
•Is there any dust or dirt on
the laser emitting/receiving
window?
Is the laser emission indicator
LED lamp lit?
Wasn’t "------" displayed
immediately after turning ON
the power switch?
Wasn’t "------" displayed
immediately after switching
the measurement modes?
Wasn’t "------" displayed
immediately after shorting
the laser emission remote
control terminal?
• Isn’t the number of measure-
ments for averaging set to a
large value?
Is there condensation forming
on the equipment?
Correctly place the targets
within the effective measure-
ment area.
Correctly select the appropri-
ate measuring segment.
•Turn the power switch OFF,
then after connecting the
scanning head, turn the
power switch ON
Remove dust or dirt from the
laser emitting/receiving
window.
Short LASER REMOTE
terminals.
Wait approximately 10
seconds after turning ON the
power switch before opera-
tion.
Wait approximately 3
seconds after changing
measurement modes before
resuming operation.
Wait approximately 10
seconds after shorting this
terminal.
In the Average mode, maxi-
mum of approximately 2.6
seconds are required until
the measurement value is
displayed after the target is
positioned in the measure-
ment area.
Change the location.
p. 84
p. 28
p. 15
p. 21
ls3100_76_88.p65 08.10.1, 10:57 AM80
Seitenansicht 85
1 2 ... 81 82 83 84 85 86 87 88 89 90 91 ... 103 104