Chapter 3 Use of Common Functions
3
36
3.2.3 Monitor (analog voltage) output setting function [F]
This function is used to change the ratio of the analog voltage output value relative
to the displayed value, provided that the output voltage is between -5 and +5 V. By
performing scaling, you can change the output voltage value to the desired level. In
addition, you can select either of the following two output methods: one outputs
analog voltage according to the displayed value and the other outputs the input
value from the sensor head.
Reference: The initial state outputs the voltage equal to the displayed value.
Output scaling
The ratio of the analog voltage output relative to the displayed value can be
changed, provided that the output voltage is between -5 and +5 V. To perform
scaling, enter the desired monitor (analog voltage) output value respectively for two
desired correction points.
1. Press the
key for at least 2 seconds.
The sensor enters the common function selection mode.
2. Press the
key to make "F" flash.
3. Press the
or key to enter the monitor output setting mode.
4. Press the
key with "F-0" displayed.
The sensor enters the monitor output scaling setting state.
5.000
55
10.000
00
Point 1
Display value: 0.00
Monitor (analog voltage) output value: 0
Point 2
Display value: 10.000
Monitor (analog voltage) output value: 5
Before scaling
After scaling
Monitor (analog voltage)
output (V)
Monitor (analog voltage)
output (V)
Display
Display
Setting