4-2 Basic Settings and Advanced Settings
4-11
IB-E
4
Setting Various Functions
Set the averaging or high-pass filter for the data retrieved at the sampling rate.
Averaging
The average value is moved and averaged. If the measured values fluctuate, stable
measurements can be obtained by increasing the average number of times. When [HPF] is
set, the high-pass filter will be enabled.
List of averaging
The sampling rate, R.V. update cycle and response time differ according to the auto adjust
function settings.
Auto adjust function disabled
3. Averaging/High-Pass Filter
Item Setting range Default value
Averaging
1/2/4/8/16/64/256/1024/4096/16384/HPF 64

LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
#X'
GO
HOLD CALC
CHECK

LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
#X'
GO
HOLD CALC
CHECK
PQ
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
QWV
GO
HOLD CALC
CHECK
3. Averaging/
High-pass filter
Averaging setting
screen
4. Output mode
/
button
[MODE]
or
(page 4-11)
(page 4-11)
(page 4-13)
Averaging
Sampling rate
(ms)
R.V. refresh rate
(ms)
Response
time (ms)
1
0.08
0.08
0.16
2 0.24
4 0.4
8 0.72
16 1.36
64 5.2
256 20.56
1024 0.32 82.24
4096 1.28 328.96
16384 5.12 1315.84