4-3 Calibration Function
4-45
IB-E
4
Setting Various Functions
Setting Method (Logical Correction)
Logical correction cannot be set if “dimension mode” is selected for
“1. Measurement Mode”.
Main screen
1
Press the [MODE] and buttons for approx. 2
seconds on the main screen at the same time.
[30.CLb] is displayed on the main display (upper level).
Press the [MODE]
+
buttons for 2
seconds or more
at the same time
2
Press the
/
button to set [LoGic] and press the
button.
3
Enter “R.V. before calibration (1)” on the main display
using the
/
buttons. The “R.V. before calibration
(1)” blinks on the main display (upper level). [SET1] is
displayed on the sub display's screen (lower level).
4
Press the [SET] button.
[set1] and the “R.V. you wish to display (1)” are alternately
displayed on the sub display's screen (lower level).
[SET] button
[
set2
]
5
Enter “R.V. you wish to display (1)” on the main
display using the
/
buttons.
6
Press the [SET] button. (SET1 confirmation operation)
After [set1] blinks several times on the sub display's screen
(lower level), [set2] is displayed.
[SET] button
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD CALC
CHECK
F'(.6
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
%.D
GO
HOLD CALC
CHECK
Calibration setting
Setting value Description Default value
defLt
No correction (Default setting)
MEAS
Measured correction
LoGic
Logical correction
5'6
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD CALC
CHECK
R.V. before calibration (1)
" 1. Measurement Mode" (page 4-8)
Item Setting range Default value
R.V. before
calibration (1)
- 999.99 to 999.99 0.00
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD CALC
CHECK
R.V. you wish to display (1)
" 1. Measurement Mode" (page 4-8)
Item Setting range Default value
R.V. you wish to
display (1)
- 999.99 to 999.99 0.00