4-3 Calibration Function
4-46
4
Setting Various Functions
IB-E
[
set2
]
7
Enter “R.V. before calibration (2)” on the main display
using the
/
buttons.
The “R.V. before calibration (2)” blinks on the main display
(upper level). [set2] is displayed on the sub display's screen
(lower level).
[SET] button
[SET] button
8
Press the [SET] button.
[set2] and the “R.V. you wish to display (2)” are alternately
displayed on the sub display's screen (lower level).
9
Enter “R.V. you wish to display (2)” on the main
display using the
/
buttons.
The “R.V. you wish to display (2)” is displayed on the sub
display's screen (lower level).
button
10
Press the [SET] button. (SET2 confirmation operation)
After [set2] blinks several times on the sub display's screen
(lower level), [END] blinks.
This completes the calibration.
11
Press the
button.
The main screen is restored.
5'6
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD CALC
CHECK
R.V. before calibration (2)
" 1. Measurement Mode" (page 4-8)
Item Setting range Default value
R.V. before
calibration (2)
- 999.99 to 999.99 100.00
" 1. Measurement Mode" (page 4-8)
Item Setting range Default value
R.V. you wish to
display (2)
- 999.99 to 999.99 100.00
LASER
BANK
0
1
2
3
HI
GO
LO
R.V.
ANALOG
HI SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD CALC
CHECK
R.V. you wish to display (2)