Seitenansicht 99
3-42
3
LK-G-M-NO3-E
3 Function Settings
Setting the Output Form of the Tolerance Comparator
The output form of the tolerance comparator can be set. The following three types are
available.
Setting the output form of the tolerance comparator
1
Hold down the key, and select "
oPt on
" with the key.
Set the common settings.
2
Press the key, and select function
"
d
"
with the
key.
Set the output form of the tolerance comparator.
3
Press the key, and select
"
d-1
"
with the key.
"
Hold
"
is set in this example.
4
Press the key to register the setting. Press the
key to return to the measurement state.
Function No. Output form Function
d-0 nor al
Normal Outputs according to the tolerance comparator.
d-1 kold
HOLD Holds the output that is turned ON. Releases the holding when
the measurement value is reset.
d-2 delay
Off-delay Off-delay of 60ms is applied to the Normal output. Releases
the holding when the measurement value is reset.
OUT1
HI GO LO
HI GO LO
TIM
OUT2
TIM
OUT1
HI GO LO
HI GO LO
TIM
OUT2
TIM
OUT1
HI GO LO
HI GO LO
TIM
OUT2
TIM