3 Function Settings
Setting the Common Function
This section describes the functions common to each program No.
List of Functions and Function Nos.
* The shaded cells are set by default.
* Function "
F
" sets the number of data and accumulation rate as a sub-function.
List of Default Values and Setting Ranges
Function
Function Function No.
Reference
Function display 0 1 2 3 4 5
page
A
Sampling rate
20 µs 50 µs 100 µs 200 µs 500 µs
1000 µs
3-39
YLE
20 50 100
200 500 1000
b
Mutual interference
prevention
OFF ON 3-40
AltEr oFF on
Timing synchronization
Asynchronous
Synchronous
3-41
Syn
n-Syn[
Syn
d
Comparator output format
Normal
HOLD
Off-delay
3-42
o-For
nor al
xoLd delay
E
Strobe time 2 ms 5 ms 10 ms 20 ms 3-43
S-t E
251020
F
Data storage
OFF OUT1 OUT2
OUT 1 , 2
3-44
d-Str oFF oUt1 oUt2 oUt1.2
Refer to "Accumulating the Measurement Value in the Memory (Data Storage Function)"
(page 3-54).
Function
Item Settable range Default value
Remarks
A
Sampling rate 20/50/100/200/500/1000 µs 200 µs
b
Mutual interference prevention
ON/OFF OFF
Timing synchronization Synchronous/asynchronous
Asynchronous
d
Comparator output format
Normal/hold/delay Normal
E
Strobe time 2/5/10/20 ms 2 ms
F
Data selection OFF/OUT1/OUT2/OUT1.2 OFF
Number of data 1 to 65536 65536
Accumulation rate 1/2/5/10/20/50/100/200/500/1000 x 20