4-12
4
LK-G-M-NO4-E
4 Input/Output Terminals
Electrical specications common to NPN type and PNP type
Analog voltage output / analog current output
Voltage output
Current output
Internal circuit
Voltage output
47
47
Current output
0V
102
Output range ±10.8V
(Measurement data: ±10.5 V)
Resolution 2 mV
Accuracy ±0.05% of F.S.
(F.S. = 20 V, accuracy with
reference to the displayed value)
Output impedance
about 100
Response
Delay Tme
about 10 µs (after the measurement
value has been updated)
Output range 3.36 to 20.64 mA
(Measurement data: 3.6 to
20.4 mA)
Resolution 3 µA
Accuracy ±0.25% of F.S.
(F.S. = 16 mA, accuracy with
reference to the displayed value)
Matching load 350 max.
Response
delay tme
about 10 µs (after the measurement
value has been updated)