Chapter 6 Normal Mode
38
6
66
66
6 [PROGRAM] Program Selection Function
The LS-5000 series can store all measurement conditions as a program. Up to 16
programs can be stored and loaded easily.
Setting Procedure
1. Select the SELECT PROGRAM NO. screen.
Press the [PROGRAM] button.
2. Select a program number.
Press the button corresponding to the program number to be loaded.
Note 1: When the program selection method is set to the rear terminals, the
program number cannot be switched on the front panel.
See p. 63.
Note 2: To switch a program number using the rear terminals, change the combi-
nation of the following four terminals and GND terminal.
ON: This terminal and the GND terminal are short-circuited.
OFF: This terminal is open.
Note 3: The system setup for the display unit (
see p. 104
) cannot be stored by
the program function.
Terminal No.
Program No. P1 P2 P3 P4
1 OFF OFF OFF OFF
2ONOFF OFF OFF
3 OFF ON OFF OFF
4ONONOFF OFF
5 OFF OFF ON OFF
6ONOFF ON OFF
7 OFF ON ON OFF
8ONONONOFF
9 OFF OFF OFF ON
10 ON OFF OFF ON
11 OFF ON OFF ON
12 ON ON OFF ON
13 OFF OFF ON ON
14 ON OFF ON ON
15 OFF ON ON ON
16 ON ON ON ON
P1 P2 P3 P4 GND
• • •
PROGRAM
1 16
LASER ON SELECT PROGRAM No.
NOR
MAL
P. 1
DISPLAY
DATA
LIMIT
SETUP
PROGRAMCALIB
OPTIONS
Selected program No.
1
5 87642 31
13 1615141210 119